Description
Infant Radiant Warmer YR02189 is an advanced medical device designed to provide optimal thermal support for newborns in critical conditions. This state-of-the-art warmer is equipped with micro-computer control for manual and skin temperature adjustments, along with separate windows to display set temperature, real temperature, and timing. Its flexible configuration allows for easy adjustment of the radiator and mattress tilt, ensuring maximum comfort for the infant. The YR02189 model includes an APGAR timer, four self-check alarms, and an RS-232 connector, enhancing both functionality and connectivity in neonatal care environments. Additionally, it provides gentle illumination through an integrated light, aiding medical professionals in observing their tiny patients closely.
Market Price
The Infant Radiant Warmer YR02189 is a high-quality device that is competitively priced within the healthcare market. Generally, similar products can be found within the price range of $1700 to $2500, offering a balance between advanced technology and affordability for healthcare institutions. For specific pricing and to gain access to exclusive offers, we invite you to request a quote or generate an automatic quote conveniently through the Kalstein Plus platform.
Frequently Asked Questions
- What is the skin sensor accuracy of the Infant Radiant Warmer YR02189? The accuracy of the skin sensor is ≤±0.3°C, providing reliable temperature readings essential for neonatal care.
- Can the warmer be adjusted for different temperature modes? Yes, it offers adjustable temperature ranges for both air mode (25.0°C~37.0°C) and high mode (37.1°C ~ 38.0°C).
- Does the warmer come with a phototherapy option? Yes, optional phototherapy lights, including LED phototherapy light, are available.
Advantages and Disadvantages
The Infant Radiant Warmer YR02189 offers numerous advantages, including precise temperature control, quick heating rates of ≤30 minutes, and excellent mattress temperature uniformity (≤2.0°C). Its micro-computer controls and self-check alarms enhance safety and usability. However, potential disadvantages may include the need for specialized training to maximize its diverse features and functions.
Product Use in the Field
In professional neonatal care settings, the Infant Radiant Warmer YR02189 is essential for maintaining stable body temperatures of newborns, particularly those born prematurely or with medical complications. Its features allow healthcare providers to monitor and adjust the conditions swiftly, ensuring the well-being of infants in incubators.
Recommendations
To ensure optimal performance, regularly calibrate the incubator’s sensors and conduct routine maintenance checks. It’s also beneficial for healthcare facilities to train their staff thoroughly on using the device efficiently, positioning it as a vital tool in neonatal intensive care units.
Features
- Micro-computer controlled skin temperature and manual adjustments
- Separate display windows for setting, real temperature, and timer
- Adjustable radiator and mattress for customized care
- Four self-check alarms for increased safety
- APGAR timer integration
- Built-in illumination light for enhanced visibility
- RS-232 connector for improved connectivity
- Optional accessories such as phototherapy lights and suction units
Technical Specifications
Model | YR02189 |
Electrical requirement | ~ 220V 50Hz or ~ 230V 60Hz |
Power consumption | ≤770VA |
Air mode temperature range | 25.0°C~37.0°C |
Over temperature alarm | <38.0°C |
High mode temperature range | 37.1°C ~ 38.0°C |
High mode temperature alarm | <39.0°C |
Temperature display range | 20.0°C – 40.0°C |
Temperature deviation alarm | ±1.0°C |
Skin sensor accuracy | ≤±0.3°C |
Mattress temperature Uniformity | ≤2.0°C |
Heating rate | ≤30min |
APGAR timing | 1 min. 5min. 10min |